INSTALL ICU SINK
1. Mount SINK onto the two HANGER STUDS. Hold SINK in place 
and temporarily install NYLOCK NUT & WASHER on one of the 
HANGER STUDS to retain SINK.
MOUNT ELECTRONICS BOX 
COMPLETE WASTE OUTLET CONNECTION
NOTE: Installation with ELECTRONIC BOX only is for cold 
or tempered water. Use optional MIXING VALVE with 
ELECTRONIC BOX when Hot and Cold supplies are 
roughed-in. See illustration below.
8
INSTALL ELECTRONICS HANGER BRACKET
1. Install the ELECTRONICS HANGER BRACKET onto remaining 
HANGER STUD. Install NYLOCK NUT & WASHER but do not 
tighten fully.
2. Remove NYLOCK NUT & WASHER from first HANGER STUD 
and install other end of ELECTRONICS HANGER BRACKET. 
Replace NYLOCK NUT & WASHER, tighten both ends. Do not over 
tighten, chinaware could become damaged.   
NYLOCK NUT
AND WASHER
HANGER STUD
COLD WATER OR
TEMPERED WATER
(3/8" SUPPLY)
HOT
MIXING VALVE
MIXING VALVE
SCREWS
MOUNTING PLATE
ELECTRONIC
BOX
ELECTRONIC BOX
COLD
1
2
1.  Connect SUPPLY NUT (1)  from SPOUT assembly to nipple on 
top of ENCLOSURE (2). Tighten with adjustable wrench to make a 
water tight connection.
CONNECT SPOUT HOSE TO ENCLOSURE
 
ELECTRONIC
BOX
1. Install “P” TRAP. Connect OFFSET TAILPIECE to “P” TRAP. 
TAILPIECE may have to be cut to fit installation. Use tubing cutter to 
cut TAILPIECE.
P-TRAP
TAILPIECE
1-1/4" OD
1-1/4" or 1-1/2" OD
WASTE OUTLET
ELECTRONICS
HANGER BRACKET
7
1. Remove COVER from ELECTRONICS BOX.
2. Place ELECTRONICS BOX onto MOUNTING PLATE. Align 
BOX with mounting holes. Attach with SCREWS provided.
3. Hang MOUNTING PLATE with ELECTRICAL BOX on 
HANGER BRACKET.