Quick Reference Guide
6
Inserting the Detector
a
 Caution
Infrared or intense ambient light (sun or halogen) may 
interfere with the base station/detector communication.
To insert a detector into a docking module, complete the follow-
ing:
1.
Activate the detector and wait until it is in normal 
operation.
2.
Ensure the alligator clip is closed and the ring is rest-
ing flat on the detector.
3.
Press the two release tabs on the docking module 
and open the lid.
If inserting the GasAlertMax XT, move the pump 
connector to the open position (red).
4.
Refer to the following sections to insert the detectors 
into the docking modules.
GasAlertClip Extreme / GasAlert Extreme
• Lower the detector (serial number face up) into the 
detector bay.
• Push forward to ensure the top of the detector connects 
with the top of the bay. 
• Close the lid and press until the release tabs click. 
• When the detector has been inserted correctly, the RUN 
LED(s) on the docking module light yellow and Unit 
Inserted
 displays on the base station LCD. The base 
station LCD displays the docking module number, and 
the type and serial number of the detector.
GasAlertMicro
• Insert the bottom of the detector (serial number face up) 
into the detector bay first and then lower the top into 
place.
• Close the lid and press until the release tabs click. 
• When the detector has been inserted correctly, the RUN 
LED(s) on the docking module light yellow and Unit 
Inserted
 displays on the base station LCD. The base 
station LCD displays the docking module number, and 
the type and serial number of the detector .
GasAlertMicro 5/PID/IR
Important: If the GasAlertMicro 5/PID/IR detector is fitted with 
a pump module, the diffusion adapter must be removed from 
the docking module. Refer to the MicroDock II Base Station 
User Manual
.
• Insert the GasAlertMicro 5/PID/IR (LCD facing up) at a 
45° angle and insert the bottom into the docking bay. 
Ensure the connector outlets on the bottom of the detec-
tor lock into place over the connector pins in the docking 
bay. 
• Close the lid and press until the release tabs click.