Tektronix TDS524A ユーザーズマニュアル
Performance Tests
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć39
These procedures check those characteristics that relate to the Main and
Delayed time base system and are listed as checked under
Delayed time base system and are listed as checked under
Warranted Char-
acteristics in Section 2, Specifications.
Check Accuracy for Long-Term Sample Rate, Delay Time,
and Delta Time Measurements
and Delta Time Measurements
Equipment Required:
One time-mark generator (Item 12) and one precision
coaxial cable (Item 4).
Prerequisites:
See page 1-15.
Procedure:
1.
Install the test hookup and preset the instrument controls:
a.
Hook up the test-signal source: Connect, through a 50
W
precision
coaxial cable, the time-mark output of a time-mark generator to CH 1.
Set the output of the generator for 10 ms markers. See Figure 1-12.
Set the output of the generator for 10 ms markers. See Figure 1-12.
Time-Mark
Generator
Output
50
W
Coaxial Cables
Figure 1-12: Initial Test Hookup
b.
Initialize the oscilloscope:
H
Press save/recall SETUP.
H
Press the main-menu button Recall Factory Setup.
H
Press the side-menu button OK Confirm Factory Init.
c.
Modify the initialized front-panel control settings:
H
Set the vertical SCALE to 500 mV.
H
Press VERTICAL MENU; then press the main-menu button
Coupling. Press the side-menu
Coupling. Press the side-menu
W
button to 50
W
.
H
Press SET LEVEL TO 50%.
H
Use the vertical POSITION knob to center the test signal on
screen.
screen.
H
Set the horizontal SCALE of the Main time base to 1 ms.
Time Base System
Checks
Checks