Triggering the Oscilloscope
To use USB triggering
To use USB triggering
USB trigger will trigger on a Start of Packet (SOP), End of Packet (EOP) signal,
Reset Complete (RC), Enter Suspend (Suspend), or Exit Suspend (Exit Sus)
on the differential USB data lines (D+ and D-). USB Low Speed and Full Speed
are supported by this trigger.
Reset Complete (RC), Enter Suspend (Suspend), or Exit Suspend (Exit Sus)
on the differential USB data lines (D+ and D-). USB Low Speed and Full Speed
are supported by this trigger.
1 Press the
More
key in the Trigger section of the front panel, rotate the
Entry knob until
USB
is displayed in the
Trigger
softkey, then press the
Settings
softkey to display USB trigger menu.
2 Press the
Speed
softkey to select the speed of the transaction being
probed.
You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s).