4- 5
MegaZoom Concepts and Oscilloscope Operation
Oscilloscope Responsiveness
Oscilloscope Responsiveness 
An important element of an oscilloscope is how responsive it is to control 
changes.  After a control is changed, such as the time/division, if you must wait 
for the test instrument to respond, the feedback loop between the instrument 
and the user can become difficult.  You could confuse adjustments because of 
the delay between making a change to the instrument and seeing the effect of 
that change take place.
MegaZoom Technology
In MegaZoom technology, the primary task of processing acquisition memory 
into oscilloscope waveforms using proprietary display techniques is distributed 
to waveform acquisition and rendering hardware.  The processor’s primary 
responsibility is to scan the keyboard and manage I/O ports.  Because the 
dedicated waveform translator can draw waveforms quickly, without 
interrupting the processor, the processor is dedicated to the user interface 
tasks.  User input is acted upon immediately and the oscilloscope display very 
quickly reflects the user’s changes.  This results in the highly-responsive 