Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
4-17
MegaZoom Concepts and Oscilloscope Operation
Acquisition Modes
Acquisition Modes 
Remember how when using your analog oscilloscope you had to constantly 
adjust the brightness?  You did this to see the desired level of detail in a signal, 
or sometimes to see the signal at all! 
The 
INTENSITY
 knob on the Agilent 54620/40-series oscilloscopes works like an 
analog oscilloscope; it changes how the frequency of occurrence of waveform 
portions are mapped into various brightness levels.  The 
INTENSITY
 knob allows 
you to adjust the plotted waveforms to account for various signal 
characteristics, such as fast sweep speeds and low trigger rates.  Even the least 
frequent events are drawn at a dim, but visible level.  Turning the knob fully 
clockwise will make even the most infrequent waveform bright and easy to see
In the 54620/40-series oscilloscopes, you control the detail by selecting an 
acquisition mode:
• Normal
• Peak Detect
• Average
• Realtime
• Press the 
Acquire
 key to display the Acquire menu.