4-27
MegaZoom Concepts and Oscilloscope Operation
Memory Depth/Record Length
Memory Depth/Record Length
Run/Stop versus Single
When the oscilloscope is running, the trigger processing or update rate is 
optimized over memory depth.  When you press 
Single, memory depth and 
sample rate are maximized.
Single
For a single acquisition, all available memory is filled with each acquisition, 
regardless of sweep speed. To acquire data with the longest possible record 
length, press the 
Single key.
Running
When running, versus taking a single acquisition, the memory is divided in half.  
This allows the acquisition system to acquire one record while processing the 
previous acquisition, dramatically improving the number of waveforms per 
second processed by the oscilloscope.  While running, maximizing the rate at 
which waveforms are drawn on the display provides the best picture of your 
input signal.