MegaZoom Concepts and Oscilloscope Operation
Using Digital Channels to Probe Circuits
Using Digital Channels to Probe Circuits 
You may encounter problems when using the mixed-signal oscilloscope that are 
related to probing. These problems fall into two categories: probe loading and 
probe grounding.  Probe loading problems generally affect the circuit under 
test, while probe grounding problems affect the accuracy of the data to the 
measurement instrument.  The design of the probes minimizes the first problem, 
while the second is easily addressed by good probing practices.
Input Impedance 
The logic probes are passive probes, which offer high input impedance and high 
bandwidths.  They usually provide some attenuation of the signal to the 
oscilloscope, typically 20 dB.
Passive probe input impedance is generally specified in terms of a parallel 
capacitance and resistance.  The resistance is the sum of the tip resistor value 
and the input resistance of the test instrument (see the following figure).  The 
capacitance is the series combination of the tip compensating capacitor and the 
cable, plus instrument capacitance in parallel with the stray tip capacitance to 
ground. While this results in an input impedance specification that is an accurate 
model for DC and low frequencies, the high-frequency model of the probe input 
is more useful (see the following figure).  This high-frequency model takes into 
account pure tip capacitance to ground as well as series tip resistance, and the 
cable’s characteristic impedance (Z
o
).
DC and Low-Frequency Probe Equivalent Circuit
8.5 pF
100 k
W