In the oscilloscope, an 8-bit A/D converter converts the input waveform 
into an evenly-spaced series of voltage readings. These readings are 
made once for each channel, and are stored into memory as an array of 
voltages with implicit times.
Main and Delayed Sweep 
The delayed sweep display function magnifies the contents of sample 
memory—data used by the main and delayed sweep windows is acquired 
on a single acquisition; the data corresponds to the same trigger event. 
Using the delayed sweep window, you can zoom in on a portion of the 
waveform and examine it in greater detail.
The delayed sweep operation is based on MegaZoom technology. Unlike 
analog oscilloscopes, the MegaZoom oscilloscope’s delayed sweep is not 
a second acquisition, but an expansion of the same data displayed in the 
main sweep. MegaZoom technology makes this possible because of its 
deep memory.  It is possible to capture the main display at 1 ms/div, and 
redisplay the same trigger in the delayed display at 1
µ
s/div—a 1000:1 
zoom ratio.
When you enable the delayed sweep view, the oscilloscope divides the 
display into two portions. The upper part shows the waveform according 
to the main time base; the lower shows a small portion of the waveform 
magnified according to a second time base (the delayed sweep window). 
In the delayed sweep window, the horizontal time/division knob adjusts 
the magnification of the delayed sweep portion. The delayed sweep time 
base must be at least twice as fast as the main time base, and can vary 
up to the maximum sweep speed of 5 ns/div for 54620-series and 1 ns/div 
for 54640-series.