Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
5- 6
Making Measurements
To reduce the random noise on a signal
To reduce the random noise on a signal
If the signal you are applying to the oscilloscope is noisy,  you can set up the 
oscilloscope to reduce the noise on the displayed waveform.  First, you stabilize 
the displayed waveform by removing the noise from the trigger path.  Second, 
you reduce the noise on the displayed waveform.
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Remove the noise from the trigger path by turning on either 
high-frequency reject or noise reject.
High-frequency reject (HF reject) adds a low-pass filter with the 3-dB point at 
50 kHz.  You use HF reject to remove high-frequency noise, such as AM or FM 
broadcast stations, from the trigger path.
HF Reject 
3 dB down point
Pass 
Band
0 dB
50 kHz
dc