Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
5-13
Making Measurements
To use the XY horizontal mode
To use the XY horizontal mode 
The XY horizontal mode converts the oscilloscope from a volts-versus-time 
display to a volts-versus-volts display using two input channels.  Channel 1 is 
the X-axis input, channel 2 is the Y-axis input.  You can use various transducers 
so the display could show strain versus displacement, flow versus pressure, volts 
versus current, or voltage versus frequency.  This exercise shows a common use 
of the XY display mode by measuring the phase difference between two signals 
of the same frequency with the Lissajous method.
1 Connect a sine wave signal to channel 1, and a sine wave signal of the 
same frequency but out of phase to channel 2.
2 Press the 
Autoscale
 key, press the 
Main/Delayed
 key, then press the 
XY
 
softkey.
3 Center the signal on the display with the channel 1 and 2 position ( ) 
knobs.  Use the channel 1 and 2 volts/div knobs and the channel 1 and 2 
Vernier
 softkeys to expand the signal for convenient viewing.
The phase difference angle (
θ
) can be calculated using the following formula 
(assuming the amplitude is the same on both channels):
Example of centering a signal on the display 
θ
sin
A
B
--- or 
C
D
----
=
A
B
D
C
Signal must 
be centered 
in “X”