Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
5- 32
Making Measurements
To make cursor measurements
To make cursor measurements 
The following steps guide you through the front-panel 
Cursors key.  You can use 
the cursors to make custom voltage or time measurements on the signal.
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Press the 
Cursors
 key, then press the 
Mode
 softkey.
X and Y cursor information is displayed on the softkeys.  
X, 1/
X, 
Y, and  
binary and hexadecimal values are displayed on the line above the softkeys.  The 
three cursors modes are:
Normal
X, 1/
X, and 
Y values are displayed.  
X is the difference between 
the X1 and X2 cursors and 
Y is the difference between the Y1 and Y2 cursors.
Binary
 Binary logic levels are displayed directly above the softkeys for the 
current X1 and X2 cursor positions for all displayed channels.
Hex
Hexadecimal logic levels are displayed directly above the softkeys for the 
current X1 and X2 cursor positions for all displayed channels.
In hexadecimal and binary mode, a level can display as 1 (higher than trigger 
level), 0 (lower than trigger level), indeterminate state ( ), or X (don’t care).  
In binary mode, X is displayed if the channel is turned off. In hex mode, the 
channel is interpreted as a 0 if turned off.