Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
6- 9
Utilities
To set up the I/O port to use a controller
To set up the I/O port to use a controller
Use the I/O menu to make settings if you have a controller connected to the 
oscilloscope.
1 Press the 
Controller
 softkey to set the controller interface to RS-232 or 
GPIB.
GPIB will be a selection on this softkey only if you have the optional N2757A 
GPIB Interface Module connected to the oscilloscope.
The 
RS-232 configuration to the connector on the rear panel is:
Data bits - 8
Stop bits - 1
Parity - none
2 Press the 
XON DTR
 softkey to set the RS-232 handshake.
The RS-232 handshake protocol can be set to:
XON XON (transmit on)/XOFF (transmit off), a software handshake, or
DTR data terminal ready, a hardware handshake.
The default handshake is 
DTR.
3 Press the 
Baud 
softkey to set the RS-232 baud rate.
The RS-232 baud rate can be set to 9600, 19200, 38400, or 57600.
The default baud rate is 9600.
4 If you have the optional N2757A GPIB Interface Module connected to 
the oscilloscope, turn the Entry knob to set the GPIB address shown in 
the 
Address 
softkey.
The 
Address softkey is only displayed if the GPIB module is attached to the 
oscilloscope.
Turn the Entry knob to set the GPIB address of this oscilloscope. The 
oscilloscope address can be set to any address from 0 to 30. The default 
oscilloscope address is 7.
Controller 
type
RS-232 
Baud rate
Return to 
previous menu
Handshake
GPIB 
address