Utilities
To perform service functions
To perform service functions
The Service menu allows you to:
• perform User Cal on the oscilloscope
• view User Cal status
• perform instrument Self Test
• view information about your oscilloscope model number, code revision 
information, modules attached, and User Cal status.
• Press the 
Utility
 key, then press the 
Service
 softkey to display the service 
menu.
User Cal
User Cal performs an internal self-alignment routine to optimize the signal path 
in the oscilloscope. The routine uses internally generated signals to optimize 
circuits that affect channel sensitivity, offset, and trigger parameters. 
Disconnect all inputs and allow the oscilloscope to warm up before performing 
this procedure.
54640-only -  before pressing 
Start User Cal, connect the rear-panel Trig Out 
output to analog Channel 1 and Channel 2 inputs using a BNC tee adapter and 
equal length cables.
User Cal should be performed at least once a year, any time the ambient 
temperature of the oscilloscope has changed more than 10 °C since the last User 
Cal, or after any repair.
Performing User Cal will not invalidate your Certificate of Calibration. 
Successful completion of User Cal does not certify this oscilloscope with a 
National Institute of Standards and Technology (NIST) calibration.
• Press 
Start User Cal 
to start the routine
Start 
User Cal
Start 
Self Test
Return to 
previous menu
About 
Oscilloscope
User Cal 
status