Getting Started
To compensate your analog probe
To compensate your analog probe
You should compensate you analog probes to match their characteristics to the 
oscilloscope. A poorly compensated probe can introduce measurement errors. 
To compensate a probe, follow these steps:
1 Connect the probe from channel 1 to the 
Probe Comp
 signal on the lower-
right corner of the front panel.
2 Press 
Autoscale
.
3 Use a nonmetallic tool to adjust the trimmer capacitor on the probe for 
the flattest pulse possible.
comp.cdr
Perfectly compensated
Over compensated
Under compensated