Getting Started
To verify basic oscilloscope operation
To verify basic oscilloscope operation
1 Connect an oscilloscope probe to channel 1.
2 Attach the probe to the Probe Comp output on the lower-right side of 
the front panel of the oscilloscope.
Use a probe retractable hook tip so you do not need to hold the probe.
3 Press the 
Save/Recall 
key on the front panel, then press the 
Default Setup
 
softkey under the display.
The oscilloscope is now configured to its default settings.
4 Press the 
Autoscale
 key on the front panel.
You should then see a square wave with peak-to-peak amplitude of about 5 
divisions and a period of about 4 divisions as shown below.  If you do not see 
the waveform, ensure your power source is adequate, the oscilloscope is 
properly powered-on, and the probe is connected securely to the front-panel 
channel input BNC and to the Probe Comp calibration output.
Verifying Basic Oscilloscope Operation