Front-Panel Overview
Important Oscilloscope Considerations
Important Oscilloscope Considerations 
Using Single versus Run/Stop 
The oscilloscopes have a 
Single key and a Run/Stop key. When you press Run 
(key is illuminated in green), the trigger processing and screen update rate are 
optimized over the memory depth.  Single acquisitions always use the maximum 
memory available—at least twice as much memory as acquisitions captured in 
Run mode—and the scope stores at least twice as many samples. At slow sweep 
speeds, the oscilloscope operates at a higher sample rate when Single is used 
to capture an acquisition, as opposed to running, due to the increased memory 
available.
Using Auto trigger mode versus Normal trigger mode
Normal trigger mode requires a trigger to be detected before an acquisition can 
complete. In many cases, a triggered display in not needed to check signal levels 
or activity. For these applications, use Auto trigger mode. If you only want to 
acquire specific events as specified by the trigger settings, use Normal trigger 
mode. For more detailed discussion of Auto trigger mode and Normal trigger 
mode, refer to Chapter 3, “Triggering the Oscilloscope.”
Viewing signal detail with acquire mode 
Remember how you had to constantly adjust the brightness on old analog scopes 
to see a desired level of detail in a signal, or to see the signal at all?  With the 
Agilent 54620/40-series oscilloscopes, this is not necessary. The Intensity knob 
operates much like the brightness knob on your computer screen, so you should 
set it to a level that makes for comfortable viewing, given the room lighting, and 
leave it there.  Then you can control the detail by selecting an Acquire mode: 
Normal, Peak Detect, Average, or Realtime as described in the following 
paragraphs.
Normal acquire mode  Normal mode is the acquisition mode that you will 
probably use for acquiring samples most of the time.  It compresses up to 2 
million acquisition points per channel for the 54620-series and up to 4 million 
acquisition points per channel for the 54640-series into a 1,000-pixel column 
display record.
The 54620-series 200 MSa/s sampling speed specification means that samples 
are taken every 5 ns.  The 54640-series 2 GSa/s sampling speed specification 
means that samples are taken every 500 ps.  At the faster sweep speeds, the 
running display is built from many individual triggers. If you press the 
Stop key, 
and pan and zoom through the waveform by using the Horizontal and Vertical 
knobs, only the last trigger’s acquisition will be displayed.