Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
3- 4
Triggering the Oscilloscope
To select a trigger mode: Normal, Auto, Auto Level
To select a trigger mode: Normal, Auto, Auto Level
1 Press the 
Mode/Coupling
 key.
2 Press the 
Mode
 softkey, then select 
Normal
Auto
, or 
Auto Level
 mode.
Normal mode displays a waveform when the trigger conditions are met, 
otherwise the oscilloscope does not trigger and the display is not updated.
Auto mode is the same as Normal mode, except it forces the oscilloscope to 
trigger if the trigger conditions are not met.
Auto Level mode (54620-series only) works only when edge triggering on 
analog channels or external trigger.  The oscilloscope first tries to Normal 
trigger. If no trigger is found, it searches for a signal at least 10% of full scale 
on the trigger source and sets the trigger level to the 50% amplitude point.  
If there is still no signal present, the oscilloscope auto triggers.  This mode 
is useful when moving a probe from point to point on a circuit board.
Auto Level and Auto modes 
Use the auto trigger modes for signals other than low-repetitive-rate signals and 
for unknown signal levels.  To display a dc signal, you must use one of these two 
auto trigger modes since there are no edges on which to trigger.
Auto Level mode (54620-series only) is the same as Auto mode with an 
automatic trigger level adjustment.  The oscilloscope looks at the level on the 
signals, and if the trigger level is out-of-range with respect to the signal, the 
scope adjusts the trigger level back to the middle of the signal.
When you select 
Run, the oscilloscope operates by first filling the pre-trigger 
buffer. It starts searching for a trigger after the pre-trigger buffer is filled, and 
continues to flow data through this buffer while it searches for the trigger.  While 
searching for the trigger, the oscilloscope overflows the pre-trigger buffer; the 
first data put into the buffer is the first pushed out (FIFO).  When a trigger is 
found, the pre-trigger buffer will contain the events that occurred just before 
the trigger.  If no trigger is found, the oscilloscope generates a trigger and 
displays the data as though a trigger had occurred.
When you select 
Single, the oscilloscope will fill pre-trigger buffer memory, and 
continue flowing data through the pre-trigger buffer until the auto trigger 
overrides the searching and forces a trigger.  At the end of the trace, the scope 
will stop and display the results.