Triggering the Oscilloscope
To use Pattern triggering
To use Pattern triggering
The Pattern trigger identifies a trigger condition by looking for a specified 
pattern.  This pattern is a logical AND combination of the channels. Each 
channel can have a value of high (H), low (L), don’t care (X).  A rising or falling 
edge can be specified for one channel included in the pattern.
1 Press the 
Pattern
 key in the Trigger section of the front panel to display 
the Pattern trigger menu.
2 For each analog or digital channel you want to include in the desired 
pattern, press the 
Channel 
softkey to select the channel.
This is the channel source for the H, L, X, or edge condition. As you press the 
Channel softkey (or rotate the Entry knob on mixed-signal oscilloscopes), the 
channel you select is highlighted in the 
Pattern =
 line directly above the 
softkeys and in the upper-right corner of the display next to "
Pat
".  External 
trigger may also be specified as a channel in the pattern when using the 
2-channel and 4-channel oscilloscopes.
Adjust the trigger level for the selected analog channel by turning the Trigger 
Level knob. The position of the trigger level for the analog channel is indicated 
by the trigger level icon 
 at the far left side of the display when DC coupling 
is selected  Set the threshold for the selected digital channel group by pressing 
the 
D7 Thru D0 or D15 Thru D8  key then press the Threshold softkey.  The value 
of the trigger level or digital threshold is displayed in the upper-right corner of 
the display.
Trigger level 
or threshold
Selected 
channel
Pattern 
trigger
Channel 
select
Logic Low
Logic High
Don’t care
Rising or 
falling edge