Agilent Technologies 42A ユーザーズマニュアル

ページ / 266
3-29
Triggering the Oscilloscope
To use Sequence triggering
To use Sequence triggering
Sequence trigger allows you to trigger the oscilloscope after finding a sequence 
of events.  Defining a sequence trigger requires three steps:
a Define an event to find before you search for  the trigger.
The “find” event can be a pattern, an edge from a single channel, or the 
combination of a pattern and a channel edge.
b Define the trigger event.
The “trigger on” event can be a pattern, an edge from a single channel, the 
combination of a pattern and a channel edge, or the nth occurrence of an 
edge from a single channel.
c Set an optional reset event.
If you choose to define a “reset” event, the event can be a pattern, an edge 
from a single channel, the combination of a pattern and a channel edge, or a 
timeout value.
Sequence Trigger Flowchart
Find:?
No
No
No
Reset on:?
Yes
Trigger on:?
Yes
Trigger 
holdoff
Scope 
triggers
Start
Yes