P711 Echo Lock
Use this feature to select the measurement verification process.
If a material agitator (mixer) is used in the vessel monitored, set Echo Lock for maximum 
verification or material agitator to avoid agitator blade detection. Ensure the agitator is 
always ON while the SITRANS LU01 is monitoring the vessel to avoid stationary blade 
detection.
When set for max verification or material agitator, a new measurement outside of the 
Echo Lock Window (P713) must meet the sampling criterion (P712). 
For total lock, Echo Lock Window (P713) is preset to zero 0. The SITRANS LU01 
continuously searches for the best echo according to the algorithm chosen (P820). If the 
selected echo is within the window, the window is then centered about the echo. If not, 
the window widens with each successive shot until the selected echo is within the 
window. The window then returns to its normal width.
When Echo Lock is OFF , the SITRANS LU01 responds immediately to a new 
measurement as restricted by the Max Fill / Empty Rate (P700 / P701); however, 
measurement reliability is affected.
P712 Echo Lock Sampling
The sampling criterion sets the number of consecutive echoes appearing above or below 
the echo currently locked onto, that must occur before the measurements are validated 
as the new reading (for Echo Lock P711 values: 1 or 2).
Values
0
Off
1
Maximum verification
2
*
Material agitator
3
Total  lock
Related
P700 Max Fill Rate
P701 Max Empty Rate
P712 Echo Lock Sampling
P713 Echo Lock Window
P820 Algorithm
Values
Range: 1:1 to 99:99
Format: x:y
x = the number of above echoes
y = the number of below echoes
Related
P711 Echo Lock
P711 value
P712 preset value
1, max verification
 5:5
2, material agitator
 5:2