Siemens Software version syngo CT 2005C ユーザーズマニュアル

ページ / 321
50
 
Workflow Information
Test Bolus using CARE Bolus
You can use the CARE Bolus option as a “Test Bolus“.
How to do it
1.Insert a Bolus tracking via the right mouse button 
submenu prior to the spiral.
2.Insert “contrast“ from the right mouse button con-
text menu prior to the monitoring scans.
 
Hint: By inserting “contrast“ you are interrupting the 
Auto range function, and therefore an automatic 
start of the spiral is not possible!
3.Start with the Topogram.
4.Position the premonitoring scan and the spiral 
range.
5.Perform the premonitoring scan, position and 
accept the ROI.
6.Start monitoring scans and a short amount of con-
trast (20 ml/2.5 ml/sec.).
 
Hint: With starting the spiral the system is switching 
to the Trigger subtask card. The trigger line is not 
shown at this stage.
7.Now you can read the proper delay from the Trigger 
subtask card.
8. Insert the delay in the Routine subtask card and load 
the spiral.
9.Start spiral and injector with the full amount of con-
trast.