Step 2: Pull the front panel assembly out of the case.
Step 3: Proceed with Model SC2000 “Case Removal” on page 28.
4.4.5 Internal Arrangement
4.4.6 Cleaning
Turn off the instrument and unplug all connectors.  Use a soft cloth or tissue and a
mild cleaner.  Do not use liquid or aerosol cleaners. Do not allow any cleaner
inside the instrument.
4.4.7 Vehicle Power Option
The Model SC2001 is available with a vehicle power option for operation with bat-
teries and linear DC power supplies.  See “Specifications” on page 26 voltage and
power requirements..
4.4.8 Fuse Replacement
The power-line fuses of AC-powered instruments are located within the instru-
ment's power entry module on the rear panel.  Use two 2A, 250V fast-blow fuses
(p/n 029-3026-00).
NOTICE
Due to the momentary startup inrush current of the instrument’s power supply, 
the use of switching power supplies with the SC are not recommended.