4.4 Calibration Procedure Status
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 37
Display
Status...
Main...
Channel
Calib...
TAG
F1
F2
F3
F4
F5
Failures:
No
Maintenance-Requests:
No
Temperature:
20.0 C
Operation:
Ready
0.0
100.0
Range: 1
37.50
ppm CH4
0.00
50.00
Measure
Status...
Channel
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Analyzer Module Calibration --
Calibration procedure status...
Start span calibration procedure !
Expected span gas:
50.00 ppm
Span gas name:
Methane
Start zero calibration procedure !
Expected zero gas:
0.00 ppm
Valves...
Valve position:
Samplegas
Time & Date:
12:50:55 June 16, 1998
Measure
Cancel !
Back...
More...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Calibration Procedure Status --
Procedure status:
Ready
Current/expected gas flow:
Samplegas
Concentration in span gas units:
37.50 ppm
Last zero calibration:
Success
Last span calibration:
Success
Last span calibration was:
Fri 05-29-1998 13:37:23
Successful zero+span calibrated ranges:
1+2+3+4
Remaining procedure time:
0 s
Last zero calibration was:
Fri 05-29-1998 13:32:06
---------------------- Results ----------------------
⇒
⇒
Change to the
menu "Analyzer
Module Calibration"
menu "Analyzer
Module Calibration"
Press
F5
⇒
⇒
Change to the
menu "Calibration
Procedure Status"
menu "Calibration
Procedure Status"
Press
Options:
Via the F5 key you can
change to a submenu,
where you can close all
valves or set up each valve
separately with:
Via the F5 key you can
change to a submenu,
where you can close all
valves or set up each valve
separately with:
•
zero gas or
•
span gas or
•
sample gas or
•
test gas.
(See sections 4.7 to 4.11,
p. 4-47...)
p. 4-47...)
⇒
⇒
Change to the menu
page "Calibration
Deviations"
page "Calibration
Deviations"
Press
F5