Fig. 2.8 
2.12. Temperature  Coefficients 
As the ambient temperature of 
the instrument changes from 
the original calibration 
temperature, errors will be 
introduced into the output of 
the instrument.  The 
Temperature Coefficient of 
Zero describes the change in 
the output that is seen at zero 
flow.  This error is added to the 
overall output signal regardless 
of flow, but can be eliminated 
by merely adjusting the zero 
potentiometer of the flow 
meter/controller to read zero 
volts at zero flow conditions.  
The Temperature Coefficient of 
Span describes the change in 
output after the zero error is 
eliminated.  This error cannot 
be eliminated, but can be 
compensated for 
mathematically if necessary.  
The curve pictured in Figure 2.8 
shows the span error in percent 
of point as a function of 
temperature assuming 23
0
C is 
the calibration temperature.