eutech-instruments cyberscan con 11 conductivitytds meter ユーザーズマニュアル

ページ / 80
Instruction Manual 
 
CON 11/110 
 
18
4.7 TDS 
Calibration 
4.7.1  Calibrating for TDS Using Conductivity Standards & adjusting TDS factor 
It is important to note that the increase in concentration of salts dissolved in solution 
increases the conductivity of that solution. This relationship varies from salt to salt 
and is roughly linear over a given range for a given salt. The TDS conversion factor is 
the number used by the meter to convert from conductivity to TDS.  
Instead of calibrating for TDS directly using TDS calibration standard solutions, you 
can have TDS calibration by using the conductivity calibration method and enter the 
appropriate TDS conversion factor into the meter.  
For more information regarding TDS Conversion Factor determination, please refer to 
page 73.  
 
4.7.2  Setting the TDS Conversion Factor 
The factory default setting for TDS conversion factor is 0.5. 
If your solution has a different TDS factor, you can improve 
calibration accuracy by setting the TDS factor prior to 
calibration. 
1.  Press MODE key to select TDS mode. 
2.  Press Setup key to enter SETUP Menu. 
3. Press 
the 
HOLD/ENTER 
key a few times to scroll 
through the sub programs until you view Program 
P1.4 
4.  Press the HOLD/ENTER key again. The upper 
display shows a value and the lower display shows 
“tdS”. 
5.  Press the MI/▲ or MR/▼ keys to select your 
calculated TDS conversion factor. 
6.  Press the HOLD/ENTER key to confirm selection 
and to return to the subgroup menu. 
7. Press the CAL/MEAS key to return to 
measurement mode.  
NOTE: The set up program P1.4 for setting the TDS 
conversion factor is only activated when set up menu is 
entered while the meter is in the TDS measurement mode. 
 
 
 
 
 
ppt
ppm
SETUP
p 1.4
ppt
ppm
SETUP
0.50
ppt
ppm
SETUP
0.75
SETUP
cof
p 1.0
SETUP
MI
MR
HOLD
ENTER
HOLD
ENTER
Only from TDS Measurement Mode
Figure 14: Setting the TDS 
Conversion Factor