eutech-instruments cyberscan con 11 conductivitytds meter ユーザーズマニュアル

ページ / 80
Instruction Manual 
 
CON 11/110 
 
70
14.2  CON 110 Factory Default Setting 
PROGRAM FUNCTION 
PARAMETER 
OPTING KEYS 
PARAMETER 
OPTIONS 
DEFAULT 
SETTING 
P 1.0 
Unit Configuration 
Parameter Title 
Only 
Parameter Title 
Only 
 
P 1.1 
Selection of READY Indicator and Auto 
HOLD Function 
MI/▲; MR/▼ 
ON ; OFF, 
HOLD 
ON 
P 1.2 
Selection of °C and °F indicator 
MI/▲; MR/▼ 
°C, °F 
°C 
P 1.3  
Selection of Automatic or Manual 
Temperature Compensation 
MI/▲; MR/▼ YES, 
nO  YES 
P 1.4 
Selection of TDS Conversion Factor 
MI/▲; MR/▼ 
0.40 to 1.00 
0.50 
P 2.0 
Viewing previous calibration data 
Parameter Title 
Only 
Parameter Title 
Only 
 
P 2.1 
First range calibration point 
Viewing Only 
Viewing Only 
_ _ _  
P 2.2 
Second range calibration point 
Viewing Only 
Viewing Only 
_ _ _  
P 2.3  
Third range calibration point 
Viewing Only 
Viewing Only 
_ _ _  
P 2.4 
Fourth range calibration point 
Viewing Only 
Viewing Only 
_ _ _  
P 2.5 
Fifth range calibration point 
Viewing Only 
Viewing Only 
_ _ _  
P3.0 
Viewing electrode data 
Parameter Title 
Only 
Parameter Title 
Only 
 
P3.1 
 
Effective cell constant for first range 
Viewing Only 
Viewing Only 
K = “1.00” 
P3.2 
 
Effective cell constant for second range  
Viewing Only 
Viewing Only 
K = “1.00” 
P3.3 
 
Effective cell constant for third range 
Viewing Only 
Viewing Only 
K = “1.00” 
P3.4 
 
Effective cell constant for fourth range 
Viewing Only 
Viewing Only 
K = “1.00” 
P3.5 
 
Effective cell constant for fifth range 
Viewing Only 
Viewing Only 
K = “1.00” 
P4.0: 
Auto Power OFF 
MI/▲; MR/▼ YES, 
nO  YES 
P5.0 
Reset to factory defaults 
Parameter Title 
Only 
Parameter Title 
Only 
 
P5.1 
Resetting of all calibration and clear all 
calibration data 
MI/▲; MR/▼ nO, 
YES  nO 
P5.2 
Resetting of all User settings to factory 
defaults condition 
MI/▲; MR/▼ nO, 
YES  nO 
P6.0 Memory 
clear 
MI/▲; MR/▼ nO, 
YES  nO 
P7.0 Temperature 
 
Parameter Title 
Only 
Parameter Title 
Only 
 
P7.1 
 
Setting of temperature coefficient 
MI/▲; MR/▼ 
0.0 to 10.0 
2.1 
P7.2 
 
Setting of normalisation temperature 
MI/▲; MR/▼ 20.0, 
25.0  25.0 
P8.0 
Selection of cell constant 
MI/▲; MR/▼ 
1.0, 10.0, 0.1 
K = “1.0” 
P9.0 Automatic 
Single Point Calibration 
Parameter Title 
Only 
Parameter Title 
Only 
 
P9.1 
Selection of  Automatic or Manual 
Calibration  
MI/▲; MR/▼ YES, 
nO  YES 
P9.2  
Selection of  Single or Multi Point 
Calibration 
MI/▲; MR/▼ YES, 
nO  YES 
NOTE: The HOLD/ENTER key is used to confirm every parameter opted. The CAL/MEAS on the other hand will 
escape the set up operation to return to the measurement mode.