Ideal nd2354 ユーザーズマニュアル

ページ / 10
Mechanical Characteristics (continued)
EM IMMUNITY  
- EN 50082-2 (1995), Industrial immunity
- EN 61000-4-2 (1995), ESD
- ENV 50140 (1993), EM Radiated field of RF
- EN 61000-4-4 (1995), EFT burst
- ENV 50141 (1993), RF common mode
- EN 61000-4-8 (1995), 50 Hz H-field
- EN 50082-1 (1997), Residential Immunity  
- EN 61000-4-5 (1995), Surges  
- EN 61000-4-11 (1994), Dips, Interruptions.    
(as shown in the test report reference number: 08077IEM.002)
SAFETY WARNINGS  
Note that the instrument should not be operated or powered up while the cover is off.  The analyzer has been designed
and tested to meet IEC 348 standard and is factory-shipped in proper conditions. 
Maintenance
Please send units in need of repair to:
IDEAL INDUSTRIES, INC.
Becker Place
Sycamore, Illinois  60178
Technical Service
Please call our toll free technical support hotline at 800-435-0705 for any questions, or check out our website at
www.testersandmeters.com
19
Technical Specifications
Supply voltage:    
Through an external power supplier set 120 VAC (+10% / -15%)  
Frequency:
50...60 Hz  
Burden:
8 VA  
Operation temperature:   0 / 50 ºC  
Measuring circuit:  
THREE PHASE, ARON    
Safety:  
Category II - 600 V, as per EN 61010
Voltage measurement     
Measuring range: 
20 to 500 V a.c.  (phase-neutral)  
20 to 866 V a.c.  (between phases) automatic scale adjustment  
Other voltages:
through suitable voltage transformers   
Frequency: 
45 to 65 Hz    
Current measurement
Measuring range: 
see available current clamps  
Current transducer ratio:  user-programmable  
Measurement units: 
automatic scale adjustment    
Built-in clock with rechargeable battery:  Date and time     
Display: 
LCD;  160 x 160 pixels   
RS-232  output: 
serial type output       
Internal memory:  
256 kbytes or 1 Mb according to the model    
Accuracy 
Voltage:
0.5 % of readout ± 2 digits  
Current 
0.5 % of readout ± 2 digits  
Active power
1.0 % of readout ± 2 digits
Reactive power
1.0 % of readout ± 2 digits
Measuring conditions to assure accuracy class:
- Errors due to voltage and current transformer not included
- Temperature range: 
5 ºC to 45 ºC
- Power factor:
0.5 to 1
- Measuring rang: 
between 5 % and 100 % 
Mechanical Characteristics
Case:
Portable case.  
Dimensions:
220 x 60 x 130 mm     
Connection terminals:
input/output terminals  
Keyboard/ display:
in frontal panel   
Weight:   
0,8 kg.    
RELEVANT STANDARDS 
EN 60664,EN 610110, EN 61036, VDE 110, UL 94    
EM EMISSION
- EN 61000-3-2 (1995), Harmonics 
- EN 61000-3-3 (1995), Fluctuations de tensión
- EN 50081-2 (1993), Industrial emission
- EN 55011 (1994): Conducted (EN 55022 - Clase B)
- EN 55011 (1994): Radiated (EN 55022 - Clase A)   
IDEAL INDUSTRIES, INC.
Sycamore, IL 60178, U.S.A.
800-304-3578 Customer Assistance
www.testersandmeters.com
ND 2354-2
Made in Spain
Warning
Take into account all installation instructions referred in 
sections INSTALLATION AND STARTUP, CONNECTION
INSTRUCTIONS and TECHNICAL SPECIFICATIONS OF THE
ANALYZER.