7.2 Temperature calibration (ATC mode only)
1) Select “ATC on” as described above in Section 7.4.1.
2) Press the ENTER key. The upper display indicates the
current temperature offset. The current measured temperature is
shown in the lower display.
3) Compare the current measured temperature on the controller
display to a thermometer known to be accurate. Note down the
correct temperature value.
4) Press the 
∆∆
 or 
 keys to scroll the lower display to match
the correct value. The upper display will now show the offset value.
You can offset temperature up to 
±
 5 
o
C.
5) Press the ENTER key to confirm your selection.
6) Continue with additional Advanced Set-up procedures, or
return to the Measurement mode by pressing the 
∆∆
 and 
 keys
(escape) simultaneously.
7.2.1 Setting manual temperature compensation
Note: This parameter is blanked out when the controller is set for
ATC operation.
For manual temperature compensation, you can set two different
temperatures: process and calibration. This allows calibration at a
temperature other than your process temperature. Example: setting
a calibration temperature of 25 
o
C lets you calibrate using standard
buffer solutions at 25 
o
C, even if your process temperature is a
different temperature.
1) Select “ATC oFF” as described above in section 7.4.1
2) Press the ENTER key. The upper display shows the current
process temperature and the lower display shows “P.
o
C” to indicate
process temperature.
3) Press the 
∆∆
 or 
 keys to adjust the process temperature
value. You can adjust the value from –9.9 to 125 
o
C.
4) After you set the process temperature value, press the
ENTER key. The upper display shows the current calibration
temperature and the lower display shows “C.
o
C” to indicate
calibration temperature.
5) Press the 
∆∆
 or 
 keys to adjust the calibration temperature value. You can adjust this value from
–9.9 to 125 
o
C.
6) Press the ENTER key to confirm your selection.
7) Continue with additional Advanced Set-up procedures, or return to Measurement mode by
pressing the 
∆∆
 and 
 keys (escape) simultaneously.