eutech-instruments ammonia gas ユーザーズマニュアル

ページ / 21
Instruction Manual 
 
Ammonia Gas-sensing Electrode
 
 
14
Glassware 
Clean glassware is essential for good measurement. Be sure to wash the glassware well with a mild 
detergent and rinse very well with distilled or deionized water. Clean glassware will drain without 
leaving water droplets behind. 
 
Electrodes 
The electrodes may be checked by using the procedure found in the section entitled 
Electrode 
Slope Check
 
1. 
Be sure to use either distilled or deionized water when following the procedures given in 
Electrode Slope Check
 
2. 
If electrode fails to response as expected, see the sections entitled 
Electrode Assembly
 
and 
Checking Membrane
.  Repeat electrode slope check. 
 
3. 
If the electrode still does not perform as described, determine whether the ammonia 
electrode inner body is working properly as directed in 
Checking the Electrode Inner 
Body. 
 
4. 
If the stability and the slope check out properly, but measurement problems persist, the 
standards may be of poor quality, the sample may contain interference or complexing 
agents, or the technique may be in error. (See 
Standard, Sample
 and 
Technique
 sections 
below.) 
 
5. 
Before replacing a "faulty" electrode, review the instruction manual and be sure to: 
- Clean and rinse the electrode thoroughly. 
- Prepare the electrode properly. 
- Use proper filling solution, buffer, and standards. 
- Measure correctly and accurately. 
- Review 
TROUBLESHOOTING HINTS. 
 
Standards 
The quality of results depends greatly upon the quality of the standards. ALWAYS prepare fresh 
standards when problems arise.  It could save hours of frustrating troubleshooting! Error may result 
from contamination of prepared standards, accuracy of dilution, quality of distilled water, or a 
mathematical error in calculating the concentrations. 
 
The best method for preparation of standards is by serial dilution. This means that an initial 
standard is diluted, using volumetric glassware, to prepare a second standard solution. The second 
is similarly diluted to prepare a third standard, and so on, until the desired range of standards has 
been prepared. 
 
Sample 
If the electrode works properly in standards but not in sample, look for possible interference, 
complexing agents, or substances which could affect response or physically damage the sensing 
electrode or the reference electrode. If possible, determine the composition of the samples and 
check for problems. See 
Sample Requirements
 and 
Interference