eutech-instruments 2-wire conductivity transmitter ユーザーズマニュアル

ページ / 2
68X256902  Rev  1  10/99
Before Calibration
Remove cover of Transmitter and ensure that electrode is correctly
connected as above. Connect Amp meter to the 4/20mA output of
Transmitter (Appendix 1).
Conductivity Transmitter Calibration
Prepare fresh Calibrating solutions.
1) 
Hold electrode in air (electrode MUST be dry) and Amp meter should
read 4mA; if not, calibration is necessary;
2) 
Locate “Zero adjustment” trimmer; Using a fine screwdriver,
slowly turn trimmer till Amp meter reads 4mA;
3) 
For greater accuracy, adjust the slope with Standard calibrating
solutions;
4) 
Rinse probe in de-ionised water and immerse in standard solution;
5) 
In 1413 
µµS solution, the output is 5.1304 mA;
6) 
In 2764 
µµS solution, the output is 6.2112 mA;
7) 
In 12.88 mS solution, the output is 14.304 mA;
8) 
Using a fine screwdriver, slowly turn trimmer, marked ‘FSMS’, till
Amp meter reads correct mA value with respective Standard
solutions.
9) 
Remember to disconnect Amp meter and close cover.
NB: For other Standard values, expected current output is based on:
mA = {(16/20) x Conductivity value of Standard soln} + 4mA
Specifications
Model
ECCONXTR
Range
0.00 TO 20.00 mS
Accuracy
2% Full Scale
Calibration
2 pts
Input
3-wire (2-wires for Conductivity cell; 1 wire for
Thermistor)
Output
4-20 mA
Operating
Voltage
DC 12 V to 24 V
Load
100 Ohms max. for 12 V;
600 Ohms max. for 24V
Housing
77 x 77 x 28 mm, field mountable
Appendix 1
Connecting Amp Meter for Calibration
Figure 3 - Method One
Figure 4 - Method Two
10  Ordering Information
pH
Item
Cat Code
pH Transmitter
ECPHXTR
Combination pH Electrode, General, integral 5m cable
ECARTSO05B
Combination pH Electrode, w/PMP, integral 5m cable
ECARGTSO05B
Dissolved Oxygen
Item
Cat Code
Dissolved Oxygen Transmitter
ECDOXTR
General Purpose D.O. probe, 0.5 – 40 ppm, Industrial,
integral 5m cable
ECDOGEN
General Purpose D.O. probe, 0 – 10  ppm, Industrial,
integral 5m cable
ECDOTPll
ORP
Item
Cat Code
ORP Transmitter
ECORPXTR
Combination Platinum ORP Electrode, w/PMP,
integral 5m cable
ECHTPTTSO05B
Combination Gold ORP Electrode, w/PMP, integral
5m cable
ECHTAUTTSO05B
Conductivity
Item
Cat Code
Conductivity Transmitter
ECCONXTR
2-Pin SS,k=1.0, integral 3m cable
ECCONSEN46
Temperature
Item
Cat Code
Temperature Transmitter
ECTEMPXTR
Pt 100 probe, SS316
ECPT10001M or
ECPT10005M
Power Supply
Item
Cat Code
120V Power Supply
EC-120-XTR
220V Power Supply