eutech-instruments waterproof c401 colorimeter ユーザーズマニュアル
Measurement: pH
C401/301/201/102/101
19
2.6
pH (5.9-8.2 pH units)
•
For water and wastewater (Phenol Red Method)
1.
Switch the meter on by pressing the ON/OFF key.
2.
Press the MODE key repeatedly until [PH] is displayed.
3. Press the READ/ENTER key to confirm the test parameter. The
meter will read [STbY] indicating it is in stand-by mode, waiting for
either blanking or measurement.
either blanking or measurement.
READ
ENTER
4. Select a clean and dry vial, un-screw the cap and fill with sample
water to the line on the vial. Replace the cap, ensuring it is screwed
on finger tight to the vial.
on finger tight to the vial.
5.
Ensure the outside of the vial is clean, dry and absent of any marks
or finger-prints. Gently place the vial fully into the sample well of the
instrument, whilst aligning the ▼ mark on the vial with the ▲ mark on
the meter.
or finger-prints. Gently place the vial fully into the sample well of the
instrument, whilst aligning the ▼ mark on the vial with the ▲ mark on
the meter.
6. Press the ZERO key to blank the meter to that vial; the meter will
read [ZERo] whilst this is taking place.
Note: The meter will store the last zero value it took between tests, even if it
turns off or if the batteries are removed.
turns off or if the batteries are removed.
Blanking
Sample
Sample
ZERO
pH
7. Remove the vial from the meter and un-screw the cap. Select a
bottle of pH Indicator (Phenol Red) reagent. Unscrew the cap and
holding the bottle vertically; gently squeeze 20 drops of reagent into
the contents into the vial. Replace the cap onto the reagent bottle.
Take extra care not to over-tighten the cap.
holding the bottle vertically; gently squeeze 20 drops of reagent into
the contents into the vial. Replace the cap onto the reagent bottle.
Take extra care not to over-tighten the cap.
8.
Replace the cap onto the vial, ensuring it is screwed on finger tight.
Invert the vial several times to mix.
Invert the vial several times to mix.