MEDTRUM TECHNOLOGIES Inc. JN001 ユーザーズマニュアル
29
Prepare your Patch Pump for use
When you are done practicing and ready to use your pump with insulin, you must make sure the time and date are
correct on your pump. You must also program your settings as instructed by your healthcare professional.
You will need these items before you begin:
-
Vial of rapid-acting U-100 insulin
Warning: NEVER use insulin that is cloudy, it may be old or inactive. Failure to use rapid-acting U-100 insulin, or
using insulin that is expired or inactive, may lead to hyperglycemia or diabetic ketoacidosis (DKA).
-
An unopened Reservoir Patch
-
Alcohol prep swab
Warning: If you are a first-time Insulin Management System user, your healthcare professional will guide you
through the steps for initializing and applying your first Reservoir Patch. Do NOT attempt to apply or use a Reservoir
Patch until you have been trained by your healthcare professional. Use of the System with inadequate training or
improper setup could put your health and safety at risk.
-
A 2mL disposable sterilized syringe and a fill needle with a protective cap
Warning: Medtrum offers the 2mL disposable sterilized fill syringe with the system, which matches the fill port of the
Reservoir Patch. Other fill syringes which can meet the following specifications are also permitted. Use the fill syringe
carefully and dispose it with needle capped to a puncture-proof container according to local waste disposal
regulations.
Specifications of fill syringe:
y
Reservoir volume: 2 mL
y
Needle size: 28G
y
Needle length: 8.5 mm
y
Make sure the syringe with fill needle has been properly sterilized.