  Reliability 
– This feature is 
similar to Rate but is based 
on a percent of accuracy.  
With reliability set to 99%, 
the test would end when the 
actual moisture loss was 
within 99% of the predicted 
result.  A value may be set 
that would give exceptional 
results when the actual 
result is within 90 to 99 
percent of the predicted 
result.  This ending criterion 
is less dependent on actual moisture loss than Rate
 
 
Temp Then … – These features are used when rate or predict criteria alone 
would end the test after the release of free or bound moisture but before the full 
temperature is reached.  The MAX 5000XL first checks to see that the 
temperature is within two (2) degrees of the test temperature before checking 
for the other criteria as previously described.  This can be helpful for ashing, 
which takes a while to reach the final temperature, and testing ultra low 
moisture samples, which don’t readily release moisture at the beginning of the 
test. 
 
Sample Size Options 
This feature enters a target sample 
size and the minimum and maximum 
sample range.  Sample size ranges 
can be programmed from 0.1 gram to 
99.9 grams.  The total difference in 
values between size and window must 
be between 0 grams minimum and 100 
grams maximum. 
  Sample Size - The sample size 
should be consistent for all tests 
on a particular product in order 
to get the best standard 
deviation possible.  High volatile 
materials, such as coatings, typically require smaller sample sizes (one to three 
grams).  Low moisture materials, such as plastic resins, typically require larger 
sample sizes.  The instrument is optimized for low moisture testing with a 30g 
load.  Ashing typically requires a smaller sample size to reduce the residue that 
is emitted during testing.  The instrument is optimized for ashing with a 4g load. 
 Contact your technical sales representative for appropriate sample sizes. 
  Sample Window - The number entered sets the allowable tolerance of the 
sample to be loaded.