Arizona MAX-5000XL ユーザーズマニュアル

ページ / 92
AZI Customer Service 800-528-7411 or 602-470-1414  
Page 36 of 92 
PRINT Memory Start
 
This command prints the stored parameters.  All of the numbered Memory Start Sets 
are printed, giving the values of the parameters shown below in the print out example. 
 
              COMPUTRAC MAX 5000XL MEMORY START REPORT 
  COMPANY: AZI Preferred Customer                             UNIT #: 500061 
----------------------------------------------------------------------------- 
    # ID               TST  IDL  HST  SSZ  SWN  T/I  TOT  EC  MODE 
----------------------------------------------------------------------------- 
    1 WEIGHT           105   25   25  8.0  1.0   I   No   Rate 1.0000%/Minute 
    2 TARTRATE         150   80   25  8.0  1.0   I   No   Rate 0.0250%/Minute 
    3 LACTOSE MONOHYDR 135   80   25  4.0  0.5   I   No   Rate 0.1000%/Minute 
    4*MILK MOISTURE    210   50   25  5.0  0.5   I   No   Rate 0.1000%/Minute 
    5 MILK ASH         575   25   25  5.0  0.5   T   No   Rate 0.0500%/Minute 
    6 DRIFT            200   80   25 50.0  1.0   I   No   Time  20min 
    7 Memory Start 7   160   25   25  8.0  1.0   I   No   Rate 0.1000%/Minute 
----------------------------------------------------------------------------- 
  
  128 Memory Start     160   25   25  8.0  1.0   I   No   Rate 0.1000%/Minute 
  129 Memory Start     160   25   25  8.0  1.0   I   No   Rate 0.1000%/Minute 
 
  250 Memory Start     160   25   25  8.0  1.0   I   No   Rate 0.1000%/Minute
 
 
Page #0X 
 
 
NOTE: If all locations have been entered, a full printout of the 
Memory Starts will take five pages to print all 250 sets, otherwise 
only those Memory Starts that have been entered will be printed.   
 
The column headings are abbreviations for the following: 
Identification Number of the Memory Start Set 
ID 
Identification Name of the Memory Start Set  (an * in the list 
indicates the set is linked to the next Memory Set) 
TST 
Programmed Test Temperature 
IDL 
Programmed Idle Temperature 
HST 
Programmed HISTART Temperature  
SSZ 
Sample Size 
SWN 
Sample Window 
T/I 
Test selection using Total or Individual test result 
TOT 
Test calculation based on total weight 
EC 
Ending Criteria 
MODE 
Details for selected ending criteria