Perform Calibration 
This selection starts the automatic 
process of calibration, and verification 
if selected, of the MAX 5000XL when 
the Temperature Calibration Interface 
(TCI) is connected to the unit.  A 
detailed procedure begins on page 47. 
The calibration uses an NIST traceable 
device to measure the temperature at 
the sample level and compare that to 
the temperature measured by the 
probe (temperature sensor) in the 
heater.  Any difference between the 
two readings is then supplied to the 
MAX 5000XL control circuits to adjust 
heater current to provide the correct temperature.  The verification process monitors 
the corrected probe temperature and the TCI temperature, and reports as an error any 
difference between the two.  The systematic process is explained later in this section. 
 
Perform Verification 
The Perform Verification selection 
allows a temperature verification to be 
performed at any time, as long as the 
TCI is connected to the unit.