Arizona MAX-5000XL ユーザーズマニュアル

ページ / 92
AZI Customer Service 800-528-7411 or 602-470-1414  
Page 58 of 92 
5.6. 
Parameter Expert (Optional Feature) 
A Parameter Expert option can be 
purchased for the MAX 5000XL.  The 
Parameter Expert (PE) assists a user 
through the development of a memory 
start.  PE queries the user about his 
reference method, sets up a memory 
start, and walks the user through a series 
of tests to determine the highest 
acceptable test temperature.  To speed 
development, tests are stopped if it looks 
like they will not end in a reasonable time; 
a higher test temperature will be tried 
instead.  In the end, PE programs the 
memory start, which produces a fast and 
accurate test.  PE works on most samples up to 300°C.  For ashing, PE sets up a memory 
start and runs a single test to determine the proper ending rate (if possible).  Ashing may 
require more complex parameter development.  Typical PE development time is < 2 
hours for products < 25% moisture. 
PE can be exited and reentered, and PE remembers the previous tests, so development 
will continue were it was previously stopped. 
 
Develop From Scratch 
The Parameter Expert is informative and guides the user through the steps. 
  NAME 
– This is used for the new 
memory start name. 
  REF METHOD TEMP 
– Enter the 
temperature used by the reference 
method.  If temp > 300, then PE 
assumes ashing mode and sets up 
a memory start and runs a single 
test to determine the proper 
ending rate (if possible). 
  APPLICATION 
– Enter the 
application type, i.e. food, plastics, 
etc. 
  RESULT TYPE 
– Enter the result 
type output. 
  TARGET RESULT 
– Enter the reference method result. 
  FILTER PAPER 
– Liquids and creams may require filter paper to produce 
repeatable results.  Because filter paper is expensive, try developing with “No” filter 
paper first.