11.  TROUBLE DIAGNOSIS 
  
SYMPTOM
 
DIAGNOSIS
 
REMEDY 
Final result is always 
zero or near zero 
Temperature too low 
Raise temperature, 
No heat 
See No Heat symptom later in 
table 
Sample too small 
Increase sample size, 
Final results vary too 
much 
Test temperature incorrect 
Determine 
temperature/ending criteria to 
use or call AZI at 
800-528-7411 or 
602-470-1414 for assistance 
Ending criteria incorrect 
Sample size too small 
Increase sample size, 
Samples not uniform 
Grind sample, 
see page 65 
Sample particles too large 
Blank display or very 
hard to see 
Contrast controls not set 
Adjust contrast, see page 16 
No power 
Check power cord 
connections, see page 14 
Microprocessor failure 
Turn unit OFF and back ON.  
Call AZI at 800-528-7411 or 
602-470-1414 for assistance 
Backlight connector 
disconnected 
Call AZI at 800-528-7411 or 
602-470-1414 for assistance 
Heat will not turn off 
Circuit board failure 
Call AZI at 800-528-7411 or 
602-470-1414 for assistance