Stratos LV/LV-T Technical Manual  111 
2.20 Timing 
Statistics 
2.20.1 Event Counter 
The event counter totals all of the sensed and paced events 
from all three channels.  With the event counter, the 
following events and event sequences can be registered 
over several years: 
•  Atrial: detection (As), detection during ARP (ARS), FFP 
(As_FFP), and the PMT window (As_PMT), pacing (Ap) 
•  Ventricular: RV sensing, RV pacing and LV pacing 
•  Ventricular extrasystoles are counted both as VES and 
as ventricular sense events. 
 
N
OTE
All event counter data are transmitted to the programmer 
and evaluated there, but not all events are displayed in detail 
on the programmer. 
 
Additionally, the Stratos CRT-Ps have a PMT counter and a 
counter for safety power-down overdrive. 
2.20.2 Event Episodes 
In contrast to the event counter, it is not the individual 
events, but rather the event sequences that are counted: 
•  As followed by Vs 
•  As followed by Vp 
•  Ap followed by Vs 
•  Ap followed by Vp 
•  Vx followed by Vx 
 
The event sequence V-V means two consecutive ventricular 
events (sensing or pacing) without a previous atrial event.