6  Stratos LV/LV-T Technical Manual   
For transurethral resection of the prostate, it is recommended 
that the cautery ground plate be placed under the buttocks or 
around the thigh, but not in the thoracic area where the current 
pathway could pass through or near the CRT-P system. 
1.5.2  Storage and Sterilization 
Storage (temperature) – Recommended storage temperature 
range is 5° to 55°C (41°-131°F).  Exposure to temperatures 
outside this range may result in CRT-P malfunction (see 
Section 7.1). 
Low Temperatures – Exposure to low temperatures (below 
0°C) may cause a false elective replacement indication to be 
present.  If this occurs, warm the device to room temperature 
and reset the ERI with magnet application (see Section 7.1). 
Handling – Do not drop.  If an unpackaged CRT-P is dropped 
onto a hard surface, return it to BIOTRONIK (see Section 7.1). 
FOR SINGLE USE ONLY - Do not re-sterilize the CRT-P or 
accessories packaged with the CRT-P, they are intended for 
one-time use.  
Device Packaging – Do not use the device if the packaging is 
wet, punctured, opened or damaged because the integrity of the 
sterile packaging may be compromised.  Return the device to 
BIOTRONIK. 
Storage (magnets) – Store the device in a clean area, away 
from magnets, kits containing magnets, and sources of 
electromagnetic interference (EMI) to avoid damage to the 
device. 
Temperature Stabilization – Allow the device to reach room 
temperature before programming or implanting the device.  
Temperature extremes may affect the initial device function. 
Use Before Date   Do not implant the device after the USE 
BEFORE DATE because the device may have reduced 
longevity.