Stratos LV/LV-T Technical Manual  141 
Lead Compatibility 
The Stratos CRT-Ps have been designed for connection with a 
bipolar lead in the atrium and two bipolar leads in the ventricle.  
All connections are IS-1 compatible.  Appropriate adapters (e.g., 
A1-A) should be fitted when using leads with a different 
connection. 
N
OTE
Connecting systems with a 3.2 mm configuration that do not 
expressly claim to agree with the IS-1 dimensions generally 
have to be regarded as incompatible with IS-1 connectors 
and can only be used with BIOTRONIK products together 
with an appropriate adapter.  For questions regarding lead-
device compatibility, consult your BIOTRONIK 
representative. 
In case of device replacement, ensure that the existing lead 
connector and leads are not damaged. 
 
C
AUTION
 
Lead / CRT-P Compatibility – Because of the numerous 
available 3.2-mm configurations (e.g., the IS-1 and VS-1 
standards), lead/ CRT-P compatibility should be confirmed 
with the CRT-P and/or lead manufacturer prior to the 
implantation of the system. 
IS-1, wherever stated in this manual, refers to the international 
standard, whereby leads and generators from different 
manufacturers are assured a basic fit. [Reference ISO 5841-
3:1992(E)].