148  Stratos LV/LV-T Technical Manual   
Table 32
 shows the mean
*
 expected time intervals (in months) 
from ERI to EOS at standard program for Stratos CRT-Ps.  All 
service intervals, including the above-cited nominal longevity, 
are based on the battery discharge behavior and the hybrid 
circuit properties including current consumption and replacement 
indicator.  The statistical calculations are based on 500 ohm lead 
impedances, 100% pacing, and data supplied by the battery 
manufacturer. 
Table 32: Remaining Expected Service time (ERI to EOS) 
Pacing Program 
Stratos LV, (months) 
Standard
, Mean 
Program with high pulse energy
 8 
 
                                                      
 
*
 50% of all pacemakers reach or exceed the given value 
 Standard: DDDR, 60 ppm, 3.6 V, 0.4 ms 
 High: DDDR, 90 ppm, 4.8 V, 1.0 ms