Mintek Pacemaker LV-T ユーザーズマニュアル

ページ / 172
152  Stratos LV/LV-T Technical Manual   
Dynamic AV Delay (Dual chamber only)  
low; medium; high; individual; fixed 
AV Delay Values (Dual chamber modes only) 
15… (5)…300 
AV Hysteresis 
Off; 10… (10)…100 ms 
Repetitive AV Hysteresis 
Off; 1… (1)…10 
Scan AV Hysteresis 
Off; 1… (1)…10 
Sense Compensation  
Off; -10… (-10)…-120 ms 
Safety AV Delay 
100 ms 
Ventricular Blanking Time 
30… (5)…70 ms 
Magnet effect 
Automatic; Auto; asynchronous; synchronous 
Asynchronous Magnet Effect: paces at 90 ppm. 
Automatic Magnet Effect: 10 cycles at 90 ppm asynchronous; 
thereafter synchronous with the programmed basic rate  
Synchronous Magnet Effect: synchronous with programmed 
basic rate  
High Rate Protection 
205…215 ppm 
Pulse Amplitude  
0.2... (0.1)...3.6... (0.1)...6.2; 7.2 V  
(3 channels separately programmable)