82  Stratos LV/LV-T Technical Manual   
 
Figure 5.  Ventricular blanking time and safety AV  
2.4 
Pacing and Sensing Functions 
2.4.1  Pulse Amplitude and Pulse Width 
The pulse amplitude and pulse width can be independently 
programmed for all three channels of the Stratos CRT-Ps. 
The programmed pulse amplitude determines the voltage 
applied to the heart during each pacing pulse.  The pulse 
amplitude is independently programmable for the atrial and 
ventricular channels up to 7.2 volts.  The pulse amplitude 
remains consistent throughout the service life of the CRT-Ps.  
The pacing safety margin is therefore not reduced by a decrease 
in the CRT-P's battery voltage.