Agilent Technologies 53151A ユーザーズマニュアル

ページ / 142
Chapter 1  Getting Started
Operating the Counter
Operating Guide
1-27
1
NOTE
As the graph shows, the Counter never computes power-correction values 
for loss above the zero axis. Conversely, corrections are never computed 
for gain below the zero axis. Once the correction value reaches the zero 
axis, no further corrections are applied.
When the Counter interpolates for frequencies that are above or below the 
range of frequencies specified in the currently selected profile, it never 
computes a value that would fall on the opposite side of the zero axis from 
the closest specified frequency. In other words, if there are two or more 
data points that contain loss values, the Counter never computes or 
applies a correction that would be indicative of gain. Conversely, if there 
are two or more data points that contain gain values (negative loss values), 
the Counter never applies a correction that would be indicative of loss. 
Since there can be no further change in the loss or gain values once the 
zero axis is reached, no power corrections are applied when the input 
frequency reaches or passes a point in the profile that intersects the zero 
axis. Effectively, the Counter computes only loss-correction values or 
gain-correction values—never both within the same profile.
When you enter values in power-correction data points and then exit the 
data-point display (using either the left-arrow key or the Enter key), the 
Counter immediately sorts all of the data-points into order by the 
frequency values. Therefore, if you enter a pair of values in a data point, 
exit the data-point display, and then immediately return to the display for 
that data point, you may see different values than the ones you just 
entered. The values you entered may now be contained in a differently 
numbered data point in the same profile, if they were previously entered 
out of order by frequency.
Increasing Profile Accuracy
To increase the accuracy of a power-correction profile, you can add data 
points between the existing data points in the profile (if less than 10 data 
points are in use), thus bringing the data points closer together and 
shortening the straight-line segments. If all ten data points are in use, 
you can add data points by using two or more profiles for a single test 
configuration. If you do this, you must select the profile you need (from the 
Menu) for the range of frequencies being measured. This means that you 
must select a different profile whenever the measured frequency moves 
outside the range of frequencies defined by the data points in the 
currently selected profile.