Chapter 2  Operating Your Frequency Counter
Using Power Correction
Operating Guide
2-31
2
Power Correction Theory of Operation
When the Counter interpolates between data points to determine the 
amount of correction to apply to the current measurement, it computes the 
correction based on a straight line plotted between the frequency values in 
the two closest data points. Therefore, a graph of a power-correction 
profile would show a “curve” that consists of two to nine straight-line 
segments, rather than a true curve, as shown below.
NOTE
When the frequency value of a signal is beyond either end of the curve 
specified by the values in the currently selected profile’s data points, 
the Counter computes a correction by extending the straight line defined 
in the last two data points in the profile.
As the graph above shows, the Counter never computes power-correction 
values for loss above the zero axis. Conversely, corrections are never 
computed for gain below the zero axis. When the Counter interpolates for 
frequencies that are above or below the range of frequencies specified in 
the currently selected profile, it never computes a value that would fall on 
the opposite side of the zero axis from the closest specified frequency. 
In other words, if there are two or more data points that contain loss 
values, the Counter never computes or applies a correction that would be 
indicative of gain. Conversely, if there are two or more data points that 
contain gain values (negative loss values), the Counter never applies a 
correction that would be indicative of loss. 