Philips MP60/70/80/90 사용자 설명서

다운로드
페이지 414
285
28
28
Using Timers
With the Timers function you can set timers to notify you when a specific time period has expired. 
CAUTION
Do not use the timers to schedule critical patient-related tasks.
Viewing Timers
You can view currently active timers in the 
Timers
 window or directly on the Main Screen (if a timer 
has been substituted for a numeric).
To open the 
Timers
 window:
Select the 
Timers
 SmartKey
or
Select 
Main Setup -> Timers
The maximum number of timers which can be active depends on your monitor model.
For each timer with a defined run time a progress bar is shown with the run time for that timer. Next 
to the progress bar the elapsed time or the remaining time is shown, depending on whether the timer is 
counting 
Up
 or 
Down
. On the left, the label of the timer and a symbol indicating the type of expiry 
notification is shown. 
Notification
When any timer expires (except a No Limit timer), the color changes to red and a message appears in 
the status line on the Main Screen. Be aware that if a timer has been configured to display in red, the 
color will not change when the timer expires.
Additional notification is also available and can be pre-configured to:
Alarm An INOP alarm indicates the expiry of the timer 
Sound  A single tone indicates the expiry of the timer
No Sound No additional notification
This additional notification setting is made in the monitor’s Configuration Mode and is associated 
with the timer label. The volume of the tone can also be set in Configuration Mode