Testing and Moving to Production
Op
eratio
eratio
n
App
e
ndi
ces
Mai
n
tenan
c
e and
T
roub
le
sh
oo
tin
g
oo
tin
g
Com
m
is
si
on
in
in
g
7.7
Write-protecting the configuration
To prevent unintended changes to the transmitter configuration, you can write-protect the
configuration.
configuration.
To write-protect the configuration using ProLink II:
1. Click
ProLink > Configuration > Device
.
2. Enable write-protection.
To write-protect the configuration using the Communicator:
1. Select
Diag/Service > Perform Diagnostic Action
.
2. Select
Enable Write Protect
.
To write-protect the configuration using the display:
1. Enter the display menu system.
2. Enter the off-line maintenance menu.
3. Select the Config menu and scroll to
Lock
.
4. Enable the Lock option.
Note: To disable write-protection, see Section 3.5.1.