Troubleshooting Tips
12-8
Troubleshooting
There is a dashed (----)
line on the display
instead of an ECG.
line on the display
instead of an ECG.
ECG data is not being acquired.
z
Check the patient cable is connected.
z
Check the pads or electrodes are prop-
erly applied.
erly applied.
z
Check that the desired lead is selected.
The HeartStart XLT
does not appear to be
functioning properly.
does not appear to be
functioning properly.
z
The battery is low.
z
There is a system failure.
z
Change the battery.
z
Take the device out of use and call for
service.
service.
The displayed time is
incorrect.
incorrect.
The time was not correctly set in the con-
figuration.
figuration.
Set the time in the General Settings menu
of the Configuration Mode.
of the Configuration Mode.
The printed date is
incorrect.
incorrect.
The date was not correctly set in the con-
figuration.
figuration.
Set the time in the General Settings menu
of the Configuration Mode.
of the Configuration Mode.
Table 12-3 Troubleshooting Tips (Continued)
Situation
Cause
Possible Solution