PREPARATION
Remove the protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT.
This is normal with electrodes and they will disappear when rinsed
with water.
During transport tiny bubbles of air may form inside the glass bulb.
The electrode cannot function properly under these conditions. These
bubbles can be removed by "shaking down" the electrode as you
would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in HI70300 or
HI80300 Storage Solution for at least one hour.
For refillable electrodes:
If  the fill solution (electrolyte) is less than 1 cm (½") below the fill
hole, add HI7082 or HI8082 3,5M KCl Electrolyte Solution for
double junction or HI7071 or HI8071 3,5M KCl+AgCl Electrolyte
Solution for single junction electrodes.
For a faster response unscrew the fill hole screw during measurements.
ELECTRODE CONDITIONING
AND MAINTENANCE
Reference
Filling Hole
Reference
Filling Hole
Sensitive
Wire
Reference
Wire
Reference
Junction
Glass
Bulb
Glass
Bulb
Reference
Junction
Reference
Wire
Sensitive
Wire
Glass Body
pH Electrode
Plastic Body
pH Electrode